DIAGNOdent Pen

DIAGNOdent|Pen|Operation hygiene_header_DIAGNOdent-pen - KaVo Dental

How to use the KaVo DIAGNOdent pen

DIAGNOdent|Pen|Operation pen_works - KaVo Dental1. Clean tooth surfaces preferrably with an air polishing device (e.g. PROPHYflex) to completely remove plaque, stains and calculus from fissure areas.
2.
Dry the tooth.
3.
Perform clinical examination (scanning of tooth surface.)
4.
Diagnose and evaluate quantitative measurement of DIAGNOdent pen

Possible courses of action: 
  • Chart readings and monitor over time.
  • Initiate prophylactic treatment (fluoridation, etc.) and monitor success.
  • Restore and check for residual caries.




Therapy information according to KaVo DIAGNOdent pen

Display value: Therapy:
0 - 14 No special measures.
15 - 20 Usual prophylactic measures.
21 - 30 More intensive prophylaxis or restoration: indication is dependent on

  •  caries activity.
  •  caries risk.
  •  recall interval, etc.
from 30 Restoration and more intensive prophylaxis.




KaVo Dental
11729 Fruehauf Drive
Charlotte NC 28273
888-ASK-KAVO
888-275-5286
Fax 704-927-0816